Introduction
PanAnalytical equipment can generate big number of unique warnings which leads to massive table of warnings. Consequences are failed exports and problems to load warnings in Performance Viewer. We need to delete these warnings and stop driver from logging them.
Deletion of warnings
This is done by running a SQL script which finds all UAIV0 equipment and deletes all warnings for these equipment.
Script is attached to this page (on the top right side).
Rename file to .sql.
Open in SQL Server Management Studio
Execute.
Check if something was deleted.
Equipment configuration
Turn off logging of warnings for UAIV0.
Set Message mask to 4.
System > QCX Equipment tool > Specific > Edit
Stop/Start the driver!